RemplirTM
DESCRIPTION
RemplirTM is a sterile, implantable, biocompatible, resorbable, collagen membrane intended for use in the management of peripheral nerve injuries. RemplirTM is a single-ply sheet of porcine-derived, non-crosslinked collagen. Remplir is supplied as a single device in a double PETG/Tyvek blister pack in a labelled cardboard box. Remplir is available in 4 size variants for the convenience of clinicians (15mm x 20mm, 20mm x 30mm, 30mm x 40mm, 40mm x 50mm). All size variants are identical in composition and function.
INDICATIONS FOR USE
RemplirTM is indicated for the management of peripheral nerve injuries.
PROPERTIES
RemplirTM is designed to protect the injured nerve following repair and provide a favorable environment for nerve regeneration to occur.
RemplirTM is soft and pliable with a smooth outer surface and rough inner surface, providing excellent handling characteristics.
RemplirTM is wettable and may be sutured in place, if required.
RemplirTM can be cut to the desired size and wrapped around the injured nerve in either its wet or dry form.
DIRECTIONS FOR USE
These directions are intended as a general guide and are not intended to supersede professional clinical judgement.- Follow standard procedures for nerve exposure and mobilization. Perform neurolysis and/or tensionless coaptation of nerve ends if required. If present, the gap between nerve ends should not exceed 20mm.
- Trim RemplirTM to desired size, if required.
- Hydrate if desired with sterile saline or other physiological solution. Device is full hydrated when uniformly translucent.
- Wrap RemplirTM around the site of the nerve repair, with the rough surface of the device facing the nerve.
- Fix RemplirTM in place with sutures, if required. Do not compress the nerve.
- Follow standard post-operative procedures and care. Application of RemplirTM does not modify post-operative treatment.
CONTRAINDICATIONS
The use of RemplirTM in patients with known sensitivity to porcine-derived materials or collagen is contraindicated.
ADVERSE REACTIONS
Possible complications of peripheral nerve surgery include pain, infection, swelling, inflammation, foreign body sensation and changes in nerve sensitivity (paresthesia or hyperesthesia) at the implant site.
Adverse reactions to porcine- derived collagen nerve wraps such as RemplirTM are extremely rare.
Patients should be monitored closely in the initial post-operative period to identify and address any adverse reactions that may occur.
PRECAUTIONS
Peripheral nerve repair with RemplirTM should only be undertaken by suitably qualifieded surgeons.
RemplirTM should be used with caution in infected regions.
Do not use if the product package is damaged or opened.
STORAGE AND HANDLING
Store RemplirTM in its original packaging at room temperature (15-25°C /59-77°F) in a dry place.
HOW SUPPLIED
RemplirTM is supplied in a double PETG/Tyvek blister pack contained in a labelled cardboard box.
| Size | Code |
|---|---|
| 15x20mm | ON-152 |
| 20x30mm | ON-203 |
| 30x40mm | ON-304 |
| 40x50mm | ON-405 |
MANUFACTURER
Building 191, Murdoch University
South Street, Murdoch WA 6150 Australia
SYMBOLS USED IN LABELLING
Use-by date
Batch code
Catalogue number
Date of manufacture
Sterile using irradiation
Double sterile barrier system
Do not resterilize
Do not re-use
Do not use if package is damaged and consult Instructions for use
Contains biological material of animal origin
Keep dry
Keep away from sunlight
Temperature limit
Consult instructions for use
Medical Device
Unique Device Identifier
MR Safe
Prescription only
Device size
Manufacturer